We use cookies to distinguish you from other users and to provide you with a better experience on our websites. Close this message to accept cookies or find out how to manage your cookie settings.
To save content items to your account,
please confirm that you agree to abide by our usage policies.
If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account.
Find out more about saving content to .
To save content items to your Kindle, first ensure no-reply@cambridge.org
is added to your Approved Personal Document E-mail List under your Personal Document Settings
on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part
of your Kindle email address below.
Find out more about saving to your Kindle.
Note you can select to save to either the @free.kindle.com or @kindle.com variations.
‘@free.kindle.com’ emails are free but can only be saved to your device when it is connected to wi-fi.
‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.
We report a case of a rare cochlear implant complication: the introduction of the electrode array into the superior semicircular canal, with intra-operative measurements of neural response reactions suggesting reasonable functioning of the implant.
Case report:
A two-year old patient affected by congenital, profound, sensorineural deafness underwent bilateral cochlear implantation at the ENT clinic of the ‘La Sapienza’ University of Rome. Two Clarion 90k devices were implanted, and electrophysiological and radiological checks were performed. After the introduction of the array in the right side, neural response imaging was performed, and a neural potential was found only on two apical electrodes, at a stimulation intensity of 431 clinical units. The situation differed on the left side, where neural response imaging was present at a stimulation intensity of 300 clinical units on the two electrodes tested (one apical electrode (number three), and one middle electrode (number nine)). Intra-operative radiological assessment with a transorbital plain films was performed as usual in order to assess the position of the electrodes inside the cochlea. This radiography showed the electrode array to be in the superior semicircular canal in the right ear.
Conclusion:
Intra-operative monitoring tests during cochlear implant surgery play different roles; measurement of impedances and neural response imaging can evaluate the integrity of implant electrodes and the status of the electrode–cochlea interface, but it must not be the sole way in which correct positioning of the array is confirmed. In our opinion, intra-operative radiological assessment is mandatory during cochlear implant surgery.