A ceiling-mounted far-ultraviolet-C light technology reduces methicillin-resistant Staphylococcus aureus contamination on surfaces in a simulated operating room

Published online by Cambridge University Press:  10 September 2025

Samir Memic
Affiliation:
Department of Systems Biology, Case Western Reserve University School of Medicine, Cleveland, OH, USA Research Service, Louis Stokes Cleveland VA Medical Center, Cleveland, OH, USA
Jennifer L. Cadnum
Affiliation:
Research Service, Louis Stokes Cleveland VA Medical Center, Cleveland, OH, USA
Curtis J. Donskey*
Affiliation:
Department of Medicine, Case Western Reserve University School of Medicine, Cleveland, OH, USA Geriatric Research Education and Clinical Center, Louis Stokes Cleveland VA Medical Center, Cleveland, OH, USA
*
Corresponding author: Curtis J. Donskey; Email: Curtis.Donskey@va.gov
Rights & Permissions [Opens in a new window]

Abstract

Information

Type
Letter to the Editor
Creative Commons
This is a work of the US Government and is not subject to copyright protection within the United States. Published by Cambridge University Press on behalf of The Society for Healthcare Epidemiology of America.
Copyright
© US Department of Veterans Affairs, 2025

To the Editor Cleaning and disinfection of operating room (OR) surfaces is generally recognized as an important practice, Reference Calderwood, Anderson and Bratzler1Reference Loftus, Brindeiro, Loftus, Brown, Charnin and Dexter4 although the impact on patient outcomes is uncertain. Reference Xie, Sax and Daodu5 A recent systematic scoping review concluded that the existing literature shows suboptimal compliance and inconsistent effectiveness of OR cleaning and disinfection. Reference Xie, Sax and Daodu5 Ultraviolet-C (UV-C) light room decontamination devices have been shown to reduce surface contamination when operated as an adjunct to manual cleaning and disinfection. Reference El Haddad, Ghantoji and Stibich6 However, use of these devices can require a substantial amount of time and the devices can only be operated when people are not present. Reference El Haddad, Ghantoji and Stibich6 Far-UV-C light (200–230 nm) is an alternative technology that may provide similar efficacy while being safe in occupied areas. Reference Memic, Torres-Teran, Cadnum and Donskey7 Here, we evaluated the efficacy of a commercial, ceiling-mounted far-UV-C technology for continuous adjunctive decontamination of surfaces in a simulated OR.

The Visium 1 far-UV-C technology (Lit Thinking, Orlando, Florida) that was tested uses a krypton-chloride excimer lamp (Care222, Ushio America, Cypress, California) that emits a primary wavelength of 222 nm. We tested the Clear Optics version of Visium 1 which emits light over 60-degree angles. With Clear Optics devices placed a minimum of 1.2 m (3.9 ft) apart and operating continuously at a height of ≥2.5 m (8.2 ft), far-UV-C doses are within threshold limit values proposed by the American Conference of Governmental Industrial Hygienists (ie, 161 mJ/cm2 for eyes and 479 mJ/cm2 for skin) over an 8-hour workday. Reference Memic, Torres-Teran, Cadnum and Donskey7 Users can also choose an “occupied off” mode in which the device runs continuously but turns off when motion is detected, thereby providing decontamination between cases with no far-UV-C exposure to personnel. Each device is intended to provide far-UV-C coverage for an area up to 18.6 m2 square meters (200 ft2); multiple devices can be installed to provide coverage for larger rooms. The supplementary material provides detailed information on the technology and shows results of initial laboratory testing demonstrating efficacy of a single device against methicillin-resistant Staphylococcus aureus (MRSA) with 2 and 8 h exposure times.

Testing was conducted in an 8.1 × 6.8 m (26.6 × 22.3 ft) simulated OR with a 3 m (9.8 ft) ceiling height. Six Visium 1 devices with Clear Optics were installed at a height of 2.8 m (9.2 ft) above the floor. A modification of the American Society for Testing and Materials (ASTM) standard quantitative disk carrier test method (ASTM E 2197-02) was used. 8 The test organism was a clinical MRSA strain; MRSA was chosen for testing given reports that it may be transmitted in the anesthesia workspace. Reference Munoz-Price, Bowdle and Johnston3,Reference Loftus, Brindeiro, Loftus, Brown, Charnin and Dexter4 A 10 µl inoculum containing 104–105 colony-forming units (CFU) of MRSA was spread to cover 20 mm steel disks with 5% fetal calf serum as a soil load. The inoculated disks were placed at 10 sites in the room within 3.05 m (10 ft) of the nearest lamp, including horizontal, vertical, and partially shaded sites. The disk locations included the OR patient tabletop (horizontal), patient table side (vertical), patient table bottom (partially shaded), floor beside patient table, adjustable light top, adjustable light bottom (partially shaded), sterile item table top, sterile item table side (vertical), floor beside sterile item table, and workstation in the corner of the room. Far-UV-C irradiance was measured using a radiometer (UIT2400 Handheld Light Meter for 222 nm (Ushio America, Cypress, CA). Testing was completed with exposure times of 4 and 16 hours. The supplementary material provides a picture of the simulated OR with the locations of the far-UV-C devices and the test sites.

Figure 1 shows reductions in MRSA and irradiance readings in the simulated OR with 6 Clear Optics devices. After 4 hours of exposure, MRSA was reduced by ≥2 log10 CFU at 5 of 10 sites. After 16 hours of exposure, MRSA was reduced by ≥3 log10 CFU at 6 of 10 sites and by ≥2 log10 CFU at 8 of 10 sites. The sites with the lowest levels of reduction had low irradiance readings and were partially shaded (ie, bottom of the patient table and bottom of an adjustable light) or in the corner of the room outside the area of maximal light exposure.

Figure 1. Reduction in methicillin-resistant Staphylococcus aureus (MRSA) on steel disk carriers in a simulated operating room after 4 and 16 hours of exposure to far-ultraviolet-C (UV-C) light delivered by 6 ceiling-mounted Clear Optics devices. White diamonds indicate irradiance measured at each test site with a radiometer. Error bars show standard error. Log10 reductions were calculated in comparison to untreated controls.

In summary, we demonstrated that a ceiling-mounted far-UV-C technology reduced MRSA contamination at multiple sites in a simulated OR. After 4 or 16 hours of exposure, MRSA was reduced by >2 log10 or >3 log10 at 50% or more of the test sites. Although light delivery was reduced in partially shaded areas, average reductions of 1.8 to 2.3 log10 CFU of MRSA were achieved after 16 hours of exposure. Given that multiple studies have demonstrated suboptimal cleaning and disinfection in ORs, 2,Reference Munoz-Price, Bowdle and Johnston3,Reference Xie, Sax and Daodu5 our results suggest that the far-UV-C technology could be useful as an adjunctive measure. The potential advantages of the far-UV-C technology over standard UV-C room decontamination devices include minimal personnel time requirement after installation and continuous decontamination during procedures.

Our study has some limitations. Testing was conducted in a simulated OR with inoculated steel disk carriers in a limited number of sites. Additional testing is needed to assess efficacy for real-world surfaces in ORs. However, previous studies have demonstrated that far-UV-C light is effective in reducing contamination on a variety of surfaces including portable equipment. Reference Memic, Torres-Teran, Cadnum and Donskey7,Reference Huang, Yang and Hsiao9 We did not assess ozone production. However, we have previously demonstrated that no ozone accumulates in well-ventilated (ie, 8 air changes per hour) hospital rooms with 6 Ushio Care 222 devices operating. Reference Memic, Kaple, Cadnum and Donskey10

Supplementary material

The supplementary material for this article can be found at https://doi.org/10.1017/ice.2025.10231

Acknowledgements

We thank Lit Thinking, Inc for providing the Visium technology used for testing and STERIS Corporation for use of the simulated OR.

Financial support

Supported by the Department of Veterans Affairs.

Competing interests

C.J.D has received research grants from Clorox and Ushio America. All other authors report no conflicts of interest relevant to this article.

References

Calderwood, MS, Anderson, DJ, Bratzler, DW, et al. Strategies to prevent surgical site infections in acute-care hospitals: 2022 update. Infect Control Hosp Epidemiol 2023;44:695720.10.1017/ice.2023.67CrossRefGoogle ScholarPubMed
Association of perioperative Registered Nurses (AORN). Guidelines for perioperative practice: Environmental cleaning. Association of perioperative Registered Nurses (AORN) eGuidelines, 2020 edition. Available at: https://aornguidelines.org/guidelines/content?sectionid=173715702&view=book#236401885. Accessed February 1, 2025.Google Scholar
Munoz-Price, LS, Bowdle, A, Johnston, BL, et al. Infection prevention in the operating room anesthesia work area. Infect Control Hosp Epidemiol 2019;40:117.CrossRefGoogle ScholarPubMed
Loftus, RW, Brindeiro, CT, Loftus, CP, Brown, JR, Charnin, JE, Dexter, F. Characterizing the molecular epidemiology of anaesthesia work area transmission of Staphylococcus aureus sequence type 5. J Hosp Infect 2024;143:186194.10.1016/j.jhin.2023.07.003CrossRefGoogle ScholarPubMed
Xie, A, Sax, H, Daodu, O, et al. Environmental cleaning and disinfection in the operating room: a systematic scoping review through a human factors and systems engineering lens. Infect Control Hosp Epidemiol 2024;45:880889.10.1017/ice.2023.280CrossRefGoogle ScholarPubMed
El Haddad, L, Ghantoji, SS, Stibich, M, et al. Evaluation of a pulsed xenon ultraviolet disinfection system to decrease bacterial contamination in operating rooms. BMC Infect Dis 2017;17:672.CrossRefGoogle ScholarPubMed
Memic, S, Torres-Teran, MM, Cadnum, JL, Donskey, CJ. Evaluation of a far ultraviolet-C device for decontamination of portable equipment in clinical areas. Am J Infect Control 2025;53:403406.CrossRefGoogle ScholarPubMed
ASTM International. Designation E2197. Standard Quantitative Disk Carrier Test Method for Determining Bactericidal, Virucidal, Fungicidal, Mycobactericidal, and Sporicidal Activities of Chemicals. West Conshohocken, PA: ASTM International; 2011.Google Scholar
Huang, J-R, Yang, T-W, Hsiao, Y, et al. Far-UVC light (222 nm) efficiently inactivates clinically significant antibiotic-resistant bacteria on diverse material surfaces. Microbiol Spectr 2024;12:e0425123.10.1128/spectrum.04251-23CrossRefGoogle ScholarPubMed
Memic, S, Kaple, CE, Cadnum, JL, Donskey, CJ. Do far ultraviolet-C light technologies increase ozone concentrations in healthcare facility patient rooms?. Infect Control Hosp Epidemiol 2025: Jun 16:13. doi: 10.1017/ice.2025.10207.CrossRefGoogle ScholarPubMed
Figure 0

Figure 1. Reduction in methicillin-resistant Staphylococcus aureus (MRSA) on steel disk carriers in a simulated operating room after 4 and 16 hours of exposure to far-ultraviolet-C (UV-C) light delivered by 6 ceiling-mounted Clear Optics devices. White diamonds indicate irradiance measured at each test site with a radiometer. Error bars show standard error. Log10 reductions were calculated in comparison to untreated controls.

Supplementary material: File

Memic et al. supplementary material

Memic et al. supplementary material
Download Memic et al. supplementary material(File)
File 3.4 MB