Half-kilowatt high-energy third-harmonic conversion to 50 J @ 10 Hz at 343 nm

Published online by Cambridge University Press:  07 January 2025

Jan Pilar*
Affiliation:
HiLASE Centre, Institute of Physics of the Czech Academy of Sciences, Dolni Brezany, Czech Republic
Martin Divoky
Affiliation:
HiLASE Centre, Institute of Physics of the Czech Academy of Sciences, Dolni Brezany, Czech Republic
Jonathan Phillips
Affiliation:
Central Laser Facility, STFC Rutherford Appleton Laboratory, Didcot, UK
Martin Hanus
Affiliation:
HiLASE Centre, Institute of Physics of the Czech Academy of Sciences, Dolni Brezany, Czech Republic
Petr Navratil
Affiliation:
HiLASE Centre, Institute of Physics of the Czech Academy of Sciences, Dolni Brezany, Czech Republic
Ondrej Denk
Affiliation:
HiLASE Centre, Institute of Physics of the Czech Academy of Sciences, Dolni Brezany, Czech Republic
Patricie Severova
Affiliation:
HiLASE Centre, Institute of Physics of the Czech Academy of Sciences, Dolni Brezany, Czech Republic
Tomas Paliesek
Affiliation:
HiLASE Centre, Institute of Physics of the Czech Academy of Sciences, Dolni Brezany, Czech Republic
Danielle Clarke
Affiliation:
Central Laser Facility, STFC Rutherford Appleton Laboratory, Didcot, UK
Martin Smrz
Affiliation:
HiLASE Centre, Institute of Physics of the Czech Academy of Sciences, Dolni Brezany, Czech Republic
Thomas Butcher
Affiliation:
Central Laser Facility, STFC Rutherford Appleton Laboratory, Didcot, UK
Chris Edwards
Affiliation:
Central Laser Facility, STFC Rutherford Appleton Laboratory, Didcot, UK
Tomas Mocek
Affiliation:
HiLASE Centre, Institute of Physics of the Czech Academy of Sciences, Dolni Brezany, Czech Republic
*
Correspondence to: J. Pilar, HiLASE Centre, Institute of Physics of the Czech Academy of Sciences, Za Radnici 828, 25241 Dolni Brezany, Czech Republic. Email: pilarj@fzu.cz

Abstract

We present results of frequency tripling experiments performed at the Hilase facility on a cryogenically gas cooled multi-slab ytterbium-doped yttrium aluminum garnet laser system, Bivoj/DiPOLE. The laser produces high-energy ns pulses at 10 Hz repetition rate, which are frequency doubled using a type-I phase-matched lithium triborate (LBO) crystal and consequently frequency summed using a type-II phase-matched LBO crystal. We demonstrated a stable frequency conversion to 343 nm at 50 J energy and 10 Hz repetition rate with conversion efficiency of 53%.

Type
Research Article
Creative Commons
Creative Common License - CCCreative Common License - BY
This is an Open Access article, distributed under the terms of the Creative Commons Attribution licence (https://creativecommons.org/licenses/by/4.0), which permits unrestricted re-use, distribution and reproduction, provided the original article is properly cited.
Copyright
© The Author(s), 2025. Published by Cambridge University Press in association with Chinese Laser Press

1 Introduction

With the rapid development of high-energy and high-average-power (HE-HAP) diode-pumped laser systems, harmonic frequency conversion has become a tool to broaden the application spectrum of such lasers. Recent advances with indirectly driven thermonuclear fusion bring fusion power plant closer to our reach[ Reference Abu-Shawareb1], which will require ultraviolet (UV) lasers with a high repetition rate. Such lasers will need optical components that are tested under laser-induced damage threshold (LIDT) with large aperture beams[ Reference Han, Wang, Zhou, Feng, Jia, Li, Xiang and Zheng2, Reference Natoli, Capoulade, Piombini and Bertussi3]. Other applications include annealing of silicone to improve its electrical properties for the semiconductor industry[ Reference Elliott4]. In addition, ultra-short pulses, UV micromachining[ Reference Schubert, Rastegari, Feng, Mongin, Kamer, Kasparian, Wolf, Arissian and Diels5], UV ablation[ Reference Schubert, Rastegari, Feng, Mongin, Kamer, Kasparian, Wolf, Arissian and Diels5] and UV air ionization[ Reference Schubert, Rastegari, Feng, Mongin, Kamer, Kasparian, Wolf, Arissian and Diels5] offer performance increase over infrared (IR) laser pulses.

Phillips et al. [ Reference Phillips, Banerjee, Mason, Smith, Spear, De Vido, Ertel, Butcher, Quinn, Clarke, Edwards, Hernandez-Gomez and Collier6] demonstrated 65 W of average power with energy of 65 J at 343 nm using the multi-slab laser system DiPOLE100 and harmonic conversion in lithium triborate (LBO). Rothhardt et al. [ Reference Rothhardt, Rothhardt, Müller, Klenke, Kienel, Demmler, Elsmann, Rothhardt, Limpert and Tünnermann7] demonstrated 100 W of average power and pulse energy of 28.5 μJ at 343 nm with a femtosecond fiber laser and subsequent harmonic conversion in beta barium borate (BBO) crystals. Andral et al. [ Reference Andral, Walch, Moreno, Mahieu, Produit, Lozano, Bizet, Herkommer, Moret, André, Krötz, Metzger, Michel, Mysyrowicz, Kasparian, Lupinski, Houard and Wolf8] generated average power of 120 W and pulse energy of 118 mJ using a thin-disk picosecond laser and harmonic conversion in LBO crystals. Negel et al. [ Reference Negel, Loescher, Voss, Bauer, Sutter, Killi, Ahmed and Graf9] reported third-harmonic generation (THG) in LBO with average power output of 234 W at 343 nm in 7.7 ps pulses from a thin-disk laser operating at 300 kHz.

In this paper we report on THG of 55 J at 10 Hz repetition rate using output of the Bivoj/DiPOLE laser system, which corresponds to the average power of 550 W and a more than two times increase compared to the published state-of-the-art. We reached a conversion efficiency of 60% by using square flat-top beam profile with a top-hat pulse profile and by efficient optimization of thermally induced polarization changes by a previously developed polarimetric method described in Ref. [Reference Slezák, Sawicka-Chyla, Divoký, Pilař, Smrž and Mocek10]. After thermal stabilization, the third-harmonic energy dropped to around 50 J at 10 Hz. Further increase of the energy at 343 nm was limited by the energy available at the fundamental wavelength.

2 Experiment

The setup for conversion to 515 nm was described in the previous paper[ Reference Clarke, Phillips, Divoky, Pilar, Navratil, Hanus, Severova, Denk, Paliesek, Smrz, Mason, Butcher, Edwards, Collier and Mocek11] and used a sealed mount for temperature stabilization of the second-harmonic generation (SHG) LBO crystal, which resulted in better conversion stability. The frequency doubled output at 515 nm was converted to 343 nm using residual light at the fundamental wavelength of 1030 nm and type-II phase matching in a second LBO crystal (Coherent, Inc.) placed after the first LBO crystal. The crystal had an aperture of 60 mm × 60 mm, thickness of 12 mm and cut angles θ of 50.4° and φ of 90°. Dual-band anti-reflection (DBAR) coating for 1030 and 515 nm was used on the front face to minimize reflection to 0.1% and 0.15%, respectively. Single-layer anti-reflection (SLAR) coating for 343 nm was used on the back face to minimize reflection to less than 1.5% at 343 nm. Reflection values of the SLAR coating at 1030 and 515 nm were 1% and 2.5%, respectively. The above-mentioned values of reflectivity are provided by the coatings manufacturer. The THG LBO crystal was placed in a mount provided with temperature stabilization (IB Photonics). Both LBO crystals were kept at 30°C, as it provided the highest conversion efficiency. While the SHG oven was able to provide some cooling power through its thermoelectric stabilization unit, the THG oven was using only natural convection of ambient air as the cooling mechanism.

The layout of the harmonic conversion setup is shown in Figure 1 and was described in our previous paper[ Reference Divoky, Phillips, Pilar, Hanus, Navratil, Denk, Paliesek, Severova, Clarke, Smrz, Butcher, Edwards, Collier and Mocek12]. The output from the Bivoj laser system is transferred to the harmonic conversion setup by using two Keplerian telescopes. To increase the energy fluence on the crystal, the telescope de-magnifies the laser beam size 1.56 times from 77 mm × 77 mm to 49 mm × 49 mm.

Figure 1 Schematic layout of the conversion experiment. The laser beam is coming from the laser system via the laser beam distribution system (LAS+LBDS). Single components of the setup are denoted as follows: a quarter waveplate (QWP), a half waveplate (HWP), conversion crystals (LBO), a partially reflecting sampling wedge (SW) and a beam dump (BD). Diagnostics consists of a dichroic beamsplitter (DBS), mirrors (M), lenses (L), beamsplitters (BS), an energy meter (EM), a near-field camera (C1) and a far-field camera (C2 – not present during the experiment). The layout of diagnostic lines is the same for all three wavelengths and is shown only once.

Two pairs of zeroth-order waveplates were used to compensate for the thermally induced polarization changes of the Bivoj final amplifier and the latter pair was used also to adjust polarization at the input of the LBO crystals in order to orient it parallel with the principal plane of the crystal and thus maximize the overall conversion efficiency. After the conversion crystals, the beams at 1030, 515 and 343 nm wavelengths were absorbed in a beam dump realized by a set of colored glass filters suspended in a water tank. In order to obtain a low-power sample beam for diagnostics, a dielectrically coated sampling wedge was placed in front of the beam dump with reflection values being close to 1% for the 1030, 515 and 343 nm wavelengths. The single components of the diagnostic beam were separated according to their wavelengths by a pair of dichroic mirrors. Each diagnostic line consisted of an energy meter (Gentec-EO QE25LP) and a diagnostic camera (AVT Manta 145B) to monitor the image-relayed beam profile.

The energy meters were calibrated so that they would show the energy values present at the sampler for any of the three wavelengths. The calibration coefficients for the input (1030 nm), unconverted fundamental (1030 nm) and second-harmonic (515 nm) energy meters remained constant during the whole experiment. However, the calibration coefficient for the third-harmonic frequency (343 nm) varied by more than 20% throughout the experiment. Evolution of the calibration coefficients is shown in Figure 2. We suspect the cause was the change of reflectivity of the sampling wedge due to heating by the UV laser light. Because of the 3ω energy meter coefficient uncertainty, we decided to estimate the third-harmonic energy as the difference between the input energy and the sum of energies of unconverted fundamental and second-harmonic frequencies. The calculation of the 3ω energy makes use of all known values of transmissivity of all relevant optical coatings. The calculation estimates the 3ω energy at the end of the THG LBO (before it enters the final coating) and applies the known value of coating transmission at 3ω. The values of transmissivity used in the calculation are listed in Equation (1):

(1) $$\begin{align}&{T}_{\mathrm{SHG}-1\omega}=99.0\%, \qquad {T}_{\mathrm{in}-\mathrm{THG}-1\omega}=99.5\%,\nonumber\\& {T}_{\mathrm{SHG}-\mathrm{2}\omega }=99.0\%, \qquad {T}_{\mathrm{in}-\mathrm{THG}-\mathrm{2}\omega }=99.5\%,\nonumber\\& \qquad\qquad {T}_{\mathrm{out}-\mathrm{THG}-1\omega}=99.5\%,\nonumber\\& \qquad\qquad {T}_{\mathrm{out}-\mathrm{THG}-\mathrm{2}\omega }=97.0\%,\nonumber\\& \qquad\qquad {T}_{\mathrm{out}-\mathrm{THG}-3\omega}=98.0\%,\end{align}$$

Figure 2 Temporal evolution of the energy meter’s calibration coefficients: while the calibration coefficient for input energy (1030 nm) and unconverted fundamental 1ω (1030 nm) stabilized of the order of tens of seconds, the second-harmonic 2ω (515 nm) took hundreds of seconds and the third-harmonic 3ω (343 nm) did not stabilize at all.

where T SHG-xω denotes transmission through the whole SHG oven (two windows and LBO crystal) for the given harmonics, T in-THG-xω denotes transmission through the first coating of THG LBO and T out-THG-xω denotes transmission through the second coating of the THG LBO for the given harmonics.

The input 1030 nm energy was decreased by transmission through the SHG oven and first optical surface of THG LBO crystal (98.5%). The measured residual energies (E , E ) were increased by dividing them by transmission of the second optical surface of the THG LBO crystal. These energies were subtracted from the input to obtain 3ω energy at the THG LBO end and then decreased by the second optical surface transmission at 343 nm. The calculation is shown also in Equation (2) below. Scattering and absorption losses were neglected as they should be an order of magnitude lower than losses on anti-reflection (AR) coatings:

(2) $$\begin{align}&{E}_{3\omega}={T}_{\mathrm{out}-\mathrm{THG}-3\omega}\nonumber\\&\quad \cdot\left({E}_{\mathrm{in}\mathrm{put}}\cdot {T}_{\mathrm{SHG}}\cdot {T}_{\mathrm{in}-\mathrm{THG}}-\frac{E_{1\omega }}{T_{\mathrm{out}-\mathrm{THG}-1\omega}}-\frac{E_{2\omega }}{T_{\mathrm{out}-\mathrm{THG}-2\omega}}\right),\end{align}$$

where E xω denotes measured energy of the given harmonics, T SHG-xω denotes transmission through the whole SHG oven (two windows and LBO crystal) for the given harmonics, T in-THG-xω denotes transmission through the first coating of the THG LBO and T out-THG-xω denotes transmission through the second coating of the THG LBO for the given harmonics.

We plan to investigate this phenomenon further in a follow-up work as well as comparing the performance of our in situ diagnostics to a measurement of a diagnostic beam reflected from an uncoated wedge, which should not suffer from thermally induced reflection variations.

Detailed description of the Bivoj laser can be found in Ref. [Reference Banerjee, Mason, Ertel, Phillips, De Vido, Chekhlov, Divoky, Pilar, Smith, Butcher, Lintern, Tomlinson, Shaikh, Hooker, Lucianetti, Hernandez-Gomez, Mocek, Edwards and Collier13]. The Bivoj laser generated pulses with a flat-top temporal profile with duration of 10 ns. The profile can be adjusted arbitrarily and optimized with a closed loop within a 14 ns shaping window[ Reference Banerjee, Mason, Phillips, Smith, Butcher, Spear, De Vido, Quinn, Clarke, Ertel, Hernandez-Gomez, Edwards and Collier14]. Such optimization can be done also for temporal profiles of the frequency-converted beams if needed. The spectral bandwidth of the pulses was around 200 MHz and was given by the temporal shaper as the oscillator operates with 70 kHz spectral bandwidth. However, it was demonstrated that ytterbium-doped yttrium aluminum garnet (Yb:YAG) lasers cryogenically cooled to 77 K can generate pulses with spectral bandwidth of 0.43 nm[ Reference Reagan, Baumgarten, Jankowska, Chi, Bravo, Dehne, Pedicone, Yin, Wang, Menoni and Rocca15], so seeding the Bivoj laser with a broader spectrum is also possible. By optimizing the input and output polarization of the Bivoj system[ Reference Slezák, Sawicka-Chyla, Divoký, Pilař, Smrž and Mocek10], we increased the polarization uniformity, so that around 96% of the energy was in polarization suitable for frequency conversion. The s- and p-polarization components of the Bivoj beam profile at the input to frequency conversion setup are shown in Figure 3. These results were obtained for the output energy of 92 J, while the cooling helium gas flow rate in the main amplifier head was 150 g/s and its temperature was 120 K.

Figure 3 The s- and p-component profiles of the laser beam after passage through the de-magnifying telescope with optimized polarization at the input and output of the laser system. Images were taken after the polarizer was transmitting vertical polarization with 4% of total energy (a) or horizonal polarization with 96% of total energy (b). Beam profiles at the complementary polarizations were taken under the same conditions and were normalized to the sum of both intensities. The white lines in the pictures correspond to cross-sections through the center of the beam.

After polarization optimization, the two LBO crystals were inserted into the beam path when using a low-energy beam. The input energy on the SHG LBO crystal was then sequentially increased up to 86.5 J and the phase-matching angles of both crystals were optimized to maximize the THG yield. The energy conversion data are shown in Figure 4. The energy of THG at 343 nm reached the value of 55 J and conversion efficiency of 63.5%. Conversion efficiency was calculated as the ratio between energy exiting the THG LBO crystal at the third-harmonic frequency (E ) and input energy at the fundamental frequency onto the SHG LBO crystal (E input). If we counted only the input energy in the p-polarization, which can be converted to the second harmonic by the given LBO crystal (i.e., neglecting the energy in the unusable polarization), the conversion efficiency would reach 66%. The ratio would change to count in the input energy in p-polarization only, that is, E /E input-p-pol.

Figure 4 Temporal evolution of the energy of the third-harmonic frequency 3ω together with input energy (1030 nm) and unconverted residual energy at the fundamental 1ω (1030 nm) and second-harmonic 2ω (515 nm) frequencies. Points where the crystal phase-matching angle was optimized are marked with arrows.

Figure 5 Temporal evolution of the energy of the third-harmonic frequency 3ω (343 nm) together with input energy (1030 nm) and unconverted residual energy at the fundamental 1ω (1030 nm) and second-harmonic 2ω (515 nm) frequencies after SHG oven temperature stabilization.

As shown in the previous work[ Reference Clarke, Phillips, Divoky, Pilar, Navratil, Hanus, Severova, Denk, Paliesek, Smrz, Mason, Butcher, Edwards, Collier and Mocek11], the SHG oven was able to stabilize the temperature of the SHG LBO crystal well when only the second harmonic was generated. However, after the insertion of the THG crystal and optimization of its angle, the SHG conversion efficiency started to deteriorate. We were able to recover most of the SHG and THG performance by adjusting the phase-matching angle of the SHG crystal only. Adjustments of the THG crystal angle had no effect on the resulting THG performance; if the THG crystal was misaligned and no third harmonic was generated, the feedback disappeared. The back-reflected third-harmonic beam was hitting the frame of the crystal holder and the crystal itself. Based on this observation, we believe that there was parasitic feedback of the 343 nm radiation that caused rapid heating of the SHG crystal and instability of its temperature stabilization control. While the overall temperature increase of the SHG crystal can be compensated by adjusting its phase-matching angle, the induced temperature gradients could not be compensated in such a way.

The SHG oven took more than 40 minutes to stabilize the temperature and it required continuous tuning of the phase-matching angles to keep the heat sources at a constant level. Still, the temperature stability as well as consequent energy stability was achieved in the end, as is shown in Figure 5. However, the stable output energy at 343 nm dropped to 46–48 J.

Final phase-matching angle optimization, shown in Figure 5 (around the time of 1700 s), increased output energy to almost 50 J with conversion efficiency of 53.5%. If only the convertible energy on p-polarization is taken into account, the conversion efficiency will increase to 55.5%. This final section is shown in Figure 6 in detail.

Figure 6 Temporal evolution of the energy of the third-harmonic frequency 3ω (343 nm) together with input energy (1030 nm) and unconverted residual energy at the fundamental 1ω (1030 nm) and second-harmonic 2ω (515 nm) frequencies after SHG oven temperature stabilization and fine tuning of the SHG LBO phase-matching angle. Conversion efficiency E /E input is shown in pink and is related to the scale on the right.

Figure 7 Beam profile at the third-harmonic frequency (343 nm) with energy of more than 50 J at the repetition rate of 10 Hz in the beginning of the experiment (a) and after oven temperature stabilization for 90 minutes (b). The color bar was adjusted to better show the intensity variation in the beam in the presence of hot spots that affected the normalization.

The energy stability of the input 1030 nm beam was 0.3% root mean square (RMS) and 2.0% P-t-P (peak-to-peak). The energy stability of the 343 nm beam was showing a trend during the experiment, as in the beginning it was rather high at around 6% RMS and 22% P-t-P (Figure 4, time 75–160 s), which was caused presumably by dynamic heating of the SHG LBO crystal. After thermalization of the SHG crystal and adjustment of the SHG phase-matching angle, the energy stability value settled at 1.1% RMS and 7.5% P-t-P (Figure 6).

The near-field beam profiles at 343 nm corresponding to energy of 55 J at the beginning of the experiment and 50 J obtained after 90 minutes are shown in Figure 7. They show that a temperature gradient was created inside the crystals and conversion efficiency varied over the crystal cross-section.

3 Conclusion

For the first time a high-energy THG output of 50 J at 10 Hz was achieved, which presents a four times increase in terms of average power compared with the state-of-the-art. The THG was realized by frequency doubling and consequent frequency summing of the residual fundamental and frequency doubled beams. This was done using two LBO crystals in temperature-controlled holders. The output beam profile is square super-Gaussian with satisfactory energy uniformity. Long thermalization was required due to strong optical feedback between SHG and THG crystals caused by strong reflection from AR coatings. This will be addressed in the continuation of this work. However, after the thermalization, the energy stability of THG output was satisfactory and reached 1.1% RMS.

Acknowledgements

The European Regional Development Fund and the state budget of the Czech Republic project HiLASE CoE (CZ.02.1.01/0.0/0.0/15_006/0000674), project LasApp (CZ.02.01.01/00/22_008/0004573) and the Horizon 2020 Framework Programme (H2020) (739573).

Data availability statement

The data that support the findings of this study are openly available in Zenodo repository at https://doi.org/10.5281/zenodo.12569726, Ref. [Reference Pilar and Divoky16].

References

Abu-Shawareb, H. et al., Phys. Rev. Lett. 132, 065102 (2024).CrossRefGoogle Scholar
Han, W., Wang, F., Zhou, L. D., Feng, B., Jia, H. T., Li, K. Y., Xiang, Y., and Zheng, W. G., Chinese Phys. B 21, 077901 (2012).CrossRefGoogle Scholar
Natoli, J. Y., Capoulade, J., Piombini, H., and Bertussi, B., Proc. SPIE 6720, 672016 (2007).CrossRefGoogle Scholar
Elliott, D. J., Ultraviolet Laser Technology and Applications (Elsevier, 1995).CrossRefGoogle Scholar
Schubert, E., Rastegari, A., Feng, C., Mongin, D., Kamer, B., Kasparian, J., Wolf, J. P., Arissian, L., and Diels, J. C., New J. Phys. 19, 123040 (2017).CrossRefGoogle Scholar
Phillips, J. P., Banerjee, S., Mason, P., Smith, J., Spear, J., De Vido, M., Ertel, K., Butcher, T., Quinn, G., Clarke, D., Edwards, C., Hernandez-Gomez, C., and Collier, J., Opt. Lett. 46, 1808 (2021).CrossRefGoogle Scholar
Rothhardt, J., Rothhardt, C., Müller, M., Klenke, A., Kienel, M., Demmler, S., Elsmann, T., Rothhardt, M., Limpert, J., and Tünnermann, A., Opt. Lett. 41, 1885 (2016).CrossRefGoogle Scholar
Andral, U., Walch, P., Moreno, V., Mahieu, B., Produit, T., Lozano, M., Bizet, L., Herkommer, C., Moret, M., André, Y. B., Krötz, P., Metzger, T., Michel, K., Mysyrowicz, A., Kasparian, J., Lupinski, D., Houard, A., and Wolf, J. P., Appl. Phys. B 128, 177 (2022).CrossRefGoogle Scholar
Negel, J.-P., Loescher, A., Voss, A., Bauer, D., Sutter, D., Killi, A., Ahmed, M. A., and Graf, T., Opt. Express 23, 021064 (2015).CrossRefGoogle Scholar
Slezák, O., Sawicka-Chyla, M., Divoký, M., Pilař, J., Smrž, M., and Mocek, T., Sci. Rep. 12, 18334 (2022).CrossRefGoogle Scholar
Clarke, D., Phillips, J., Divoky, M., Pilar, J., Navratil, P., Hanus, M., Severova, P., Denk, O., Paliesek, T., Smrz, M., Mason, P., Butcher, T., Edwards, C., Collier, J., and Mocek, T., Opt. Lett. 48, 6320 (2023).CrossRefGoogle Scholar
Divoky, M., Phillips, J., Pilar, J., Hanus, M., Navratil, P., Denk, O., Paliesek, T., Severova, P., Clarke, D., Smrz, M., Butcher, T., Edwards, C., Collier, J., and Mocek, T., High Power Laser Sci. Eng. 11, e65 (2023).CrossRefGoogle Scholar
Banerjee, S., Mason, P. D., Ertel, K., Phillips, P. J., De Vido, M., Chekhlov, O., Divoky, M., Pilar, J., Smith, J., Butcher, T., Lintern, A., Tomlinson, S., Shaikh, W., Hooker, C., Lucianetti, A., Hernandez-Gomez, C., Mocek, T., Edwards, C., and Collier, J. L., Opt. Lett. 41, 2089 (2016).CrossRefGoogle Scholar
Banerjee, S., Mason, P., Phillips, J., Smith, J., Butcher, T., Spear, J., De Vido, M., Quinn, G., Clarke, D., Ertel, K., Hernandez-Gomez, C., Edwards, C., and Collier, J., High Power Laser Sci. Eng. 8, e20 (2020).CrossRefGoogle Scholar
Reagan, B. A., Baumgarten, C., Jankowska, E., Chi, H., Bravo, H., Dehne, K., Pedicone, M., Yin, L., Wang, H., Menoni, C. S., and Rocca, J. J., High Power Laser Sci. Eng. 6, e11 (2018).CrossRefGoogle Scholar
Pilar, J. and Divoky, M., https://doi.org/10.5281/zenodo.12569726 (2024).CrossRefGoogle Scholar
Figure 0

Figure 1 Schematic layout of the conversion experiment. The laser beam is coming from the laser system via the laser beam distribution system (LAS+LBDS). Single components of the setup are denoted as follows: a quarter waveplate (QWP), a half waveplate (HWP), conversion crystals (LBO), a partially reflecting sampling wedge (SW) and a beam dump (BD). Diagnostics consists of a dichroic beamsplitter (DBS), mirrors (M), lenses (L), beamsplitters (BS), an energy meter (EM), a near-field camera (C1) and a far-field camera (C2 – not present during the experiment). The layout of diagnostic lines is the same for all three wavelengths and is shown only once.

Figure 1

Figure 2 Temporal evolution of the energy meter’s calibration coefficients: while the calibration coefficient for input energy (1030 nm) and unconverted fundamental 1ω (1030 nm) stabilized of the order of tens of seconds, the second-harmonic 2ω (515 nm) took hundreds of seconds and the third-harmonic 3ω (343 nm) did not stabilize at all.

Figure 2

Figure 3 The s- and p-component profiles of the laser beam after passage through the de-magnifying telescope with optimized polarization at the input and output of the laser system. Images were taken after the polarizer was transmitting vertical polarization with 4% of total energy (a) or horizonal polarization with 96% of total energy (b). Beam profiles at the complementary polarizations were taken under the same conditions and were normalized to the sum of both intensities. The white lines in the pictures correspond to cross-sections through the center of the beam.

Figure 3

Figure 4 Temporal evolution of the energy of the third-harmonic frequency 3ω together with input energy (1030 nm) and unconverted residual energy at the fundamental 1ω (1030 nm) and second-harmonic 2ω (515 nm) frequencies. Points where the crystal phase-matching angle was optimized are marked with arrows.

Figure 4

Figure 5 Temporal evolution of the energy of the third-harmonic frequency 3ω (343 nm) together with input energy (1030 nm) and unconverted residual energy at the fundamental 1ω (1030 nm) and second-harmonic 2ω (515 nm) frequencies after SHG oven temperature stabilization.

Figure 5

Figure 6 Temporal evolution of the energy of the third-harmonic frequency 3ω (343 nm) together with input energy (1030 nm) and unconverted residual energy at the fundamental 1ω (1030 nm) and second-harmonic 2ω (515 nm) frequencies after SHG oven temperature stabilization and fine tuning of the SHG LBO phase-matching angle. Conversion efficiency E/Einput is shown in pink and is related to the scale on the right.

Figure 6

Figure 7 Beam profile at the third-harmonic frequency (343 nm) with energy of more than 50 J at the repetition rate of 10 Hz in the beginning of the experiment (a) and after oven temperature stabilization for 90 minutes (b). The color bar was adjusted to better show the intensity variation in the beam in the presence of hot spots that affected the normalization.