We use cookies to distinguish you from other users and to provide you with a better experience on our websites. Close this message to accept cookies or find out how to manage your cookie settings.
To save content items to your account,
please confirm that you agree to abide by our usage policies.
If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account.
Find out more about saving content to .
To save content items to your Kindle, first ensure no-reply@cambridge.org
is added to your Approved Personal Document E-mail List under your Personal Document Settings
on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part
of your Kindle email address below.
Find out more about saving to your Kindle.
Note you can select to save to either the @free.kindle.com or @kindle.com variations.
‘@free.kindle.com’ emails are free but can only be saved to your device when it is connected to wi-fi.
‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.
Transcatheter implantation of pulmonary balloon-expandable stent-valves requires pre-stenting of the right ventricular outflow tract with large calibre stents. To increase awareness of the associated risks of this part of transcatheter pulmonary valve replacement therapy, we report potential fatal complications during the implantation of AndraStents® in the right ventricular outflow tract in six cases from five different European institutions and their management.
Method and result:
We present a retrospective case series analysis looking at the time period from 2013 to 2018. Of 127 AndraStents® implanted in the right ventricular outflow tract, in six patients, age from 13 to 71 years, a misconfiguration of the AndraStent® occurred forming a “diabolo”-configuration. During inflation of the balloon, the stent showed extreme “dog-boning”, an expansion of the stent at both ends with the middle part remaining unexpanded. This led to rupture of the balloon and loss of manoeuvrability in four patients. Out of the total six cases, in four patients the stent was eventually expanded with high-pressure balloons, and in one case the stent was surgically retrieved. In one patient, in whom a percutaneous retrieval of the embolised stent was attempted, a fatal bleeding occurred.
Conclusions:
Pre-stenting of the right ventricular outflow tract by AndraStents® can lead to misconfiguration of the stent with potentially fatal complications. Rescue strategies of misconfigured stents include stent inflation and placement with high pressure non-compliant balloons or surgical backup. Interventional retrieval measures of AndraStents® cannot be advised.