The effect of body-conforming passive wearable device with knee flexion taping on dynamic knee stability

Published online by Cambridge University Press:  26 August 2025

Sung-Jin Park
Affiliation:
Department of Mechanical Engineering, Korea Advanced Institute of Science and Technology, Daejeon, Republic of Korea
Seongok Chae
Affiliation:
Department of Mechanical Engineering, Korea Advanced Institute of Science and Technology, Daejeon, Republic of Korea
Hyung-Soon Park*
Affiliation:
Department of Mechanical Engineering, Korea Advanced Institute of Science and Technology, Daejeon, Republic of Korea
*
Corresponding author: Hyung-Soon Park; Email: hyungspark@kaist.ac.kr

Abstract

Passive wearable devices are widely used for fitness and have also become fashionable. There is increasing interest in adding functionality, such as knee stability, to these compact devices, which are more convenient for daily wear than separate devices like braces or exoskeletons. This study designed and assessed flexion taping passive wearable devices (FTPW). The design emphasized providing adequate flexion moment capacity and controlling varus/valgus movement to prevent knee injuries. In this research, 20 healthy women performed single leg drop (SLD) and step-up (SU) tests with and without muscle fatigue. Knee joint angle, muscle activation, metabolic cost, and blood flow were measured across FTPW, passive wearable devices without flexion taping (PW), and control shorts (Ctrl). In the SLD test after muscle fatigue, FTPW produced a significantly larger knee flexion angle during landing. In the SU test, before and after fatigue, knee varus angle was notably higher with FTPW. Additionally, FTPW showed reduced knee flexor fatigue, indicated by smaller median frequency shifts, and improved blood flow compared to PW. No significant differences in respiratory exchange ratio were detected among the three conditions. Overall, FTPW demonstrated strong potential to enhance knee kinematics, muscle activation, and blood flow, pointing to benefits for both performance improvement and injury prevention. By delivering focused support in a compact format, FTPW may serve as an innovative passive wearable solution that supports daily movement, comfort, and daily activities. This emphasizes the device’s promise as an alternative to bulkier knee aids, merging style and functionality effectively.

Information

Type
Research Article
Creative Commons
Creative Common License - CCCreative Common License - BY
This is an Open Access article, distributed under the terms of the Creative Commons Attribution licence (http://creativecommons.org/licenses/by/4.0), which permits unrestricted re-use, distribution and reproduction, provided the original article is properly cited.
Copyright
© The Author(s), 2025. Published by Cambridge University Press

1. Introduction

The knee joint is the most commonly injured body part when the foot contacts the ground (Ford et al., Reference Ford, Myer and Hewett2003). Especially in the case of women, injuries appear repeatedly and chronically for the following reasons: persistent and abnormal forward movement of the tibia in relation to the femur (Kernozek et al., Reference Kernozek, Torry, van Hoof, Cowley and Tanner2005; Arden and Nevitt, Reference Arden and Nevitt2006), lower value of the hamstring/quadriceps strength ratio (Holcomb et al., Reference Holcomb, Rubley, Lee and Guadagnoli2007), and large knee valgus angle (Myer et al., Reference Myer, Paterno, Ford, Quatman and Hewett2006).

In particular, previous studies reported that overly small knee flexion and excessively large knee valgus angle may be leading causes of knee injuries (Chappell et al., Reference Chappell, Yu, Kirkendall and Garrett2002; Kernozek et al., Reference Kernozek, Torry, van Hoof, Cowley and Tanner2005). Additionally, imbalanced muscle reaction forces, such as excessively high quadriceps muscle force and low hamstring muscle contraction force, can contribute to knee injuries during landing. Of particular concern is the significant gender disparity in noncontact knee injury rates (McLean et al., Reference McLean, Walker and van den Bogert2005). For instance, in sports such as basketball, soccer, team handball, and daily activity, women are reported to experience knee injuries two to seven times more frequently than men. The sports medicine community has been actively trying to change this trend, with new injury prevention strategies being actively and extensively studied (Alentorn-Geli et al., Reference Alentorn-Geli, Myer, Silvers, Samitier, Romero, Lázaro-Haro and Cugat2009).

To address these challenges, various wearable devices, such as knee braces, have been applied to increase knee stability and to prevent knee injuries (Yang and Lim, Reference Yang and Lim2014). For example, in an epidemiological study of American football players, wearing a knee brace reduced the injury rate by ~2.7 times (Rishiraj et al., Reference Rishiraj, Taunton, Lloyd-Smith, Woollard, Regan and Clement2009). It was reported that knee braces were effective in keeping the knee from excessive flexion or extension during exercise and decreasing the load applied to the knee by adjusting hamstring muscles and gastrocnemius muscle activity involved in knee flexion (Wang et al., Reference Wang, Liu, Wei, Wu and Yuan2023). Furthermore, using the knee valgus brace increases the knee’s mediolateral stability and reduces the risk factors (Ramsey et al., Reference Ramsey, Briem, Axe and Snyder-Mackler2007). However, it has been noted that individuals without knee injuries found knee braces inconvenient to carry or wear, often expressing discomfort due to tightly strapped bands.

Another approach to prevent injury is to use clinical kinesio tape made of adhesive flexible materials that do not hinder movements in daily activities and can be applied directly to the skin. However, its practical limitations, such as high cost, potential for skin irritation, and the necessity of professional application, restrict its widespread usability, particularly for nonprofessional users in daily settings. Despite these benefits, taping is not easily applied by individuals in daily life due to the high price, potential skin troubles, and the need for professional expertise for accurate application.

On the other hand, interest in passive wearable devices has been increasing recently, particularly due to their potential to integrate advanced functionalities, such as elastic taping, which aligns with the goals of enhancing both performance and injury prevention in daily activities. The functional effects that occur after wearing passive wearable devices are broadly divided into biomechanical and physiological aspects. The biomechanical effects of passive wearable devices are reduction of muscle tremors and repetitive jumping power enhancement (Kramer et al., Reference Kramer, Kümmel, Gollhofer, Armbrecht, Ritzmann, Belavy and Gruber2018). The physiological effect helps a quick recovery after exercise by increasing blood flow and reducing fatigue through muscle oxygenation (Troynikov et al., Reference Troynikov, Ashayeri, Burton, Subic, Alam and Marteau2010; Joyner and Casey, Reference Joyner and Casey2015; McNeil et al., Reference McNeil, Allen, Olympico, Shoemaker and Rice2015). Because of these effects, not only athletes but also recreational sports participants can wear passive wearable devices for improving performances.

In addition to this, there is an attempt to enhance the performance of passive wearable devices by adding elastic taping. While such clothing has the disadvantage of providing less assistance compared to assistance robots, it has the advantage of being lightweight and easy to wear. For instance, when wearing a passive wearable device with added taping to assist knee extension, the execution of sit-to-stand was performed more efficiently (Lee et al., Reference Lee, Kim and Park2020). Therefore, if the intended direction is achieved and a subtle yet continuous assisting force is applied in daily life, it is expected to have a positive effect on injury prevention in activities of daily living, similar to performance enhancement.

The injury prevention effects of wearing different types of clothing are evaluated using the single leg drop (SLD) test and the step-up (SU) test. The SLD test involves movements that expose individuals to injury risks in sports activities and can be used to assess knee joint stability. Particularly during landing, significant differences in the forces acting on the anterior cruciate ligament (ACL) are observed, depending on the knee flexion angle (Laughlin et al., Reference Laughlin, Weinhandl, Kernozek, Cobb, Keenan and O’Connor2011). The SU test, utilized in rehabilitation programs (Paz et al., Reference Paz, de Freitas Maia, Farias, Santana, Miranda, Lima and Herrington2016), measures the range of motion in knee flexion. This study will examine the effect of assisting force-induced post-fatigue on knee joint movement via the SU test.

This study introduces a novel passive wearable device design, the flexion taping passive wearable devices (FTPW), which uniquely integrates elastic taping to provide dynamic support for the knee joint. By addressing limitations of traditional braces and kinesio tape, the FTPW offers a lightweight and variable solution to enhance knee stability, particularly under fatigue-induced conditions. The FTPW’s potential applications span athletic performance enhancement, injury prevention in high-risk sports, and rehabilitation for individuals with knee instability. To evaluate its efficacy, this study investigates the effects of fatigue on kinematics, muscle activation, blood flow, and respiratory exchange ratio (RER) using the SLD and SU tests, simulating real-world conditions faced by athletes and women. To verify the effectiveness of the proposed passive wearable device, we compared the effect of wearing FTPW devices with passive wearable devices without flexion taping (PW) and loose shorts without any compression (Ctrl).

2. Materials and methods

2.1. Design of passive wearable devices with/without knee flexion taping

The PW was created as a body-conforming device that provides uniform compression. Building on this foundation, the FTPW incorporates two taping layouts to deliver task-specific support (Figure 1). An X-pattern added across the back of the knee stores elastic energy during knee extension and assists flexion by passing over the hamstrings and gastrocnemius. An I-pattern is added along the medial and lateral collateral ligament regions to limit excessive knee abduction and adduction. Thus, the FTPW retains the compression of the PW while adding functional elastic taping for flexion assistance and frontal-plane stabilization.

Figure 1. Design overview of FTPW. X-taping for knee flexion was applied to the back of the knee, and I-taping for preventing knee abduction/adduction was applied to the medial/lateral sides of the knee, respectively. To prevent slipping during motion, bands were used to anchor the garment to the ankle and waist.

X-taping (370 N/m) required elasticity to facilitate knee flexion, and for I-taping (500 N/m), higher stiffness than X-taping was set to prevent excessive knee abduction/adduction. Anchoring components were placed at the garment’s waist and ankle to avoid slipping of the edge, which decreases the amount of elastic force created by the taping. The stiffness of the taping was calculated as the ratio of the increase in length in response to applied weight. The pressure exerted by both the FTPW and PW on the pelvis and ankle was determined by comparing the wearer’s waist circumference to the device’s circumference before and after wearing, following a method described by Lee et al. (Reference Lee, Kim and Park2020). The measured pressure was 3.5 kPa at the pelvis and 2.9 kPa at the ankle for both conditions. The total weight of the FTPW and PW was 150 g.

2.2. Experimental protocol

Twenty healthy women without a history of musculoskeletal diseases and cardiovascular diseases participated in the study (age: 38.6 ± 5.3 years, weight: 63.3 ± 9.3 kg, body mass index: 23.6 ± 3.6; mean ± standard deviation [SD]). Informed consent was obtained from all subjects using the form approved by the Institutional Review Board of Korea Advanced Institute of Science and Technology (KH2021–036). An experimental protocol was designed to evaluate the effects of FTPW on kinematics, muscle activation, and RER during a fatigued state.

All subjects were tested while wearing three different types of garments: FTPW, PW, and Ctrl. To prevent carryover effects, a randomized three-way crossover design was employed. To minimize the effects of fatigue and learning between experimental sessions, the test interval was set to 1 week.

Changes in kinematics due to fatigue were evaluated using the SLD and SU tests (Figure 2). Twenty-one reflective markers were placed on lower extremity landmarks based on a Plug-in Gait marker system (Kadaba et al., Reference Kadaba, Ramakrishnan and Wootten1990). In the SLD test, subjects stood on their dominant leg on a 30 cm high box and were instructed to drop to a spot on the force plate (Advanced Mechanical Technology, Inc., MA, USA) 30 cm in front of the box without making any additional compensatory movements after landing. For the SU test, participants stepped up onto a 30 cm box with their dominant foot and then stepped down again with the same dominant foot after planting both feet at the top of the box. Both the SLD and SU tests were repeated three times before and after the fatigue protocol. To maintain the effect of fatigue, these tests were performed within 30 s of completing the fatigue protocol.

Figure 2. The experimental procedure in this study. All participants completed three trials, at a 1-week interval. SLD/SU tests were performed before and after fatigue riding the Climbmill.

The fatigue protocol involved the use of a Climbmill (Shandong Minolta Fitness Equipment Co., Ltd, Ningjin Town, China), a gradual fatigue-inducing machine that targets the lower limbs. In this study, six electromyographic (EMG) sensors were attached to vastus medialis (VM), vastus lateralis (VL), semitendinosus (ST), biceps femoris (BF), medial gastrocnemius (MGT), and lateral gastrocnemius (LGT). We used an EMG sensor covered by cushions to reduce pressure increase. Subjects wore indirect calorimetry (Cosmed K5: Cosmed, Rome, Italy) to calculate the metabolic cost during the fatigue protocol. This machine had 30 cm high steps and operated at speeds ranging from 24 to 164 steps/min. Subjects climbed at an increased speed of 20 steps/min every 3 min after the exercise began. The fatigue protocol continued until the subject’s heart rate reached 80% of maximum heart rate, which is the criterion for fatigue in this experiment. The heart rate meter employed in this study was the Polar OH1 (Polar Electro Oy, Kempele, Finland). The maximum heart rate was calculated using the formula: HRmax = 220 − age, according to the preceding paper (Mj, Reference Mj1957).

Before testing, subjects rested for a minimum of 15 min to allow their heart rates to stabilize. Once heart rates reached a stable state, blood flow velocity in the common carotid artery was measured for 10 s using an ultrasound blood flow meter. The blood flow velocity was measured using ultrasonography (SONON 300 L, Healcerion Co., Ltd, Seoul, Republic of Korea) at a sampling rate of 10 Hz. The measurement was performed three times to obtain an accurate reading.

2.3. Data analysis

For motion analysis, the VICON NEXUS system (Oxford Metric Ltd, Oxford, UK) consisting of 12 cameras runs at a sampling rate of 100 Hz. Following data collection from the SLD and SU tests, the knee joint angle, knee joint moment, knee joint force, and ground reaction force (GRF) data were processed using Visual 3D (C-motion Inc., Germantown, MD). The knee joint angle was calculated as the angle of the local coordinate system of each femur and tibia (Wu et al., Reference Wu, Siegler, Allard, Kirtley, Leardini, Rosenbaum and Witte2002). To standardize the timing of each test, the data were normalized as percentages of the task period. During the SLD test, the starting point of the motion was set at 0% and the time point when the dominant foot touched the floor was set as 25%. The remaining time after landing was adjusted to match 100% or three times the length of the landing time. During the SU test, the point when the dominant foot began to move on the floor was set to 0% and the point when the opposite foot touched the floor was set to 100%. When normalized, the motion phases could be approximately divided into the dominant foot stepping onto the box (0–25%), the opposite foot stepping onto the box (26–50%), the dominant foot stepping down to the floor (51–75%), and the opposite foot stepping down to the floor (76–100%).

For EMG analysis, the Trigno Wireless EMG system (Delsys Inc., Boston, MA, USA) was employed at a sampling rate of 2,000 Hz. A 30-s window of muscle activity at both the start and end of the fatigue protocol was used when calculating the median frequency. A 50–500 Hz bandpass filter was applied during data processing. Data from the start and end of the fatigue protocol were used to divide the power spectral density into two regions containing equal amounts of power via fast Fourier transform (Hsu et al., Reference Hsu, Tseng, Chen, Wang, Yang, Lin and Liu2020). This is the midpoint, a medium frequency used to evaluate the start and end of the fatigue protocol.

RER is calculated as the ratio of the produced CO2 volume to the consumed O2 volume. Average RER values during the final minute of the fatigue protocol were computed to discern differences among the various experimental conditions. Blood flow rate (BFR) was calculated by multiplying the cross-sectional area of the common carotid artery and the average velocity of the blood flow, as specified in a previous study (Schorer et al., Reference Schorer, Badoual, Bastide, Vandebrouck, Licker and Sage2017).

To assess participants’ subjective experiences while wearing the devices, a post-experiment subject assessment was conducted for the PW and FTPW conditions. The Ctrl condition was excluded from the survey because it provided neither compression nor structural support and served only as a baseline condition. The survey used a 5-point Likert scale to assess comfort, pressure sensation, and the general feeling of wearing the device, with a total of 14 inquiries (Chyung et al., Reference Chyung, Roberts, Swanson and Hankinson2017).

2.4. Statistical analysis

The sphericity of the data was assessed using Mauchly’s sphericity test. In cases where Mauchly’s sphericity test yielded a positive result, repeated-measures analysis of variance (ANOVA) with Tukey’s post-hoc analysis was employed to determine significant differences between experimental conditions. Differences in subjective assessments between the FTPW and PW conditions were examined with paired t-tests. Statistical analysis for median frequency difference, RER, BFR, peak vertical GRF (vGRF), and subject assessments between conditions was performed using SPSS Statistics 26 (IBM Corp., New York, USA). To determine the statistical significance of the knee joint angle, joint moment, and joint power at each time point during the SLD and SU tests, Statistical Parametric Mapping analysis was employed (Pataky et al., Reference Pataky, Robinson and Vanrenterghem2013).

3. Results

Regarding kinematic analysis, before fatigue (Figure 3), there were no significant differences in sagittal/frontal plane kinematics during the SLD test (Figure 3b, 3c). However, during the SU test, knee varus was significantly greater in FTPW than PW and Ctrl when stepping up (Figure 3d). There were no significant differences in the sagittal plane kinematics (Figure 3c). In the kinematic analysis after fatigue (Figure 4), the knee flexion angle at landing was significantly greater in FTPW than PW and Ctrl during the SLD test (Figure 4a). Additionally, during the SU test, knee varus was significantly greater in FTPW than PW and Ctrl when stepping up (Figure 4d). No significant differences were found in frontal/sagittal plane kinematics during the SLD/SU test, respectively (Figure 4b, 4c).

Figure 3. Knee joint rotation during the SLD/SU test (a, b/c, d) before the fatigue protocol. The black, blue, and magenta lines represent Ctrl, PW, and FTPW, respectively. Error bars mean standard error. The gray-shaded area corresponds to sections where the repeated-measures ANOVA results indicate a statistically significant difference (p < .05). Below each graph, bars compare group differences at each time point: Ctrl versus PW, Ctrl versus FTPW, and PW versus FTPW. The color of bars corresponds to the group with larger values.

Figure 4. Knee joint rotation during the SLD/SU test (a, b/c, d) after the fatigue protocol. The black, blue, and magenta lines represent Ctrl, PW, and FTPW, respectively. Error bars mean standard error. The gray-shaded area corresponds to sections where the repeated-measures ANOVA results indicate a statistically significant difference (p < .05). Below each graph, bars compare group differences at each time point: Ctrl versus PW, Ctrl versus FTPW, and PW versus FTPW. The color of bars corresponds to the group with larger values.

In the kinetic analysis, knee joint flexion moment (Supplementary Figure S5A), superior force acting on the knee joint (Supplementary Figure S5E), and peak vGRF (Supplementary Figure S6) were all significantly lower in the FTPW than in Ctrl.

In the kinematic analysis, dividing the subjects into groups of 10 each based on the varus knee alignment of their dominant leg’s tibiofemoral angle (low varus knee; −9.83 ± 1.51°, high varus knee; −7.21 ± 1.29°; mean ± SD), both groups demonstrated congruence in the SLD/SU test (Supplementary Figures S1–S4).

The results regarding changes in the median frequency of muscles are presented in (Figure 5). No significant differences between conditions were found in VM (F(1.62, 14.58) = 1.34, p = .29, η2 = .16, Figure 5a) and VL (F (2.00, 18.00) = .39, p = .68, η2 = .05, Figure 5b). Similarly, no significant differences between conditions were found in ST (F (1.19, 1.71) = .39, p = .59, η2 = .05, Figure 5c), MGT (F (2.00, 18.00) = .71, p = .51, η2 = .09, Figure 5e), and LGT (F (1.79, 16.11) = .25, p = .76, η2 = .03, Figure 5f). However, there was a significant difference between conditions in BF (F (1.56, 14.04) = 4.81, p = .04, η2 = .41, Figure 5d). Post-hoc analysis revealed that FTPW exhibited significantly less fatigue compared to PW (p = .032) and Ctrl (p = .002).

Figure 5. The median frequency difference of each muscle during the first and last 30 s of Climbmill exercise for the different conditions. Error bars mean standard error, and the value above the error bars is mean ± standard error. Asterisks (* and **) indicate statistically significant differences between clothing conditions (p < .05 and p < .01).

Regarding metabolic cost analysis (Figure 6), no differences were found between conditions regarding the RER (F (1.463, 13.165) = 1.34, p = .290, η2 = .13, Figure 6a). The analysis of BFR showed statistically significant differences between conditions with Ctrl, PW, and FTPW (F (2.00, 12.00) = 6.54, p = .010, η2 = .52, Figure 6b). According to the results of the post-hoc analysis, there was a statistically significant increase in BFR in FTPW compared to PW (p = .020).

Figure 6. The RER (a) and BFR (b) for each condition. Error bars mean standard error, and the value above the error bars is mean ± standard error. Asterisk (*) indicates statistically significant differences between clothing conditions (p < .05).

Regarding the subjective assessment (Table 1), paired t-tests revealed no significant differences between conditions for wearability (p = .538) or overall comfort (p = .608). Perceived compression, however, was significantly higher in the FTPW condition compared with PW (p = .022).

Table 1. Subjective assessment scores for the PW and FTPW conditions

Note. Values are presented as mean ± SD. Asterisks (*) denote statistically significant differences between clothing conditions (p < .05).

4. Discussion

The objective of this study was to investigate the knee injury prevention effects of combining passive wearable devices with knee flexion taping. The knee flexion angle at landing was significantly greater in the FTPW compared to both the PW and Ctrl in the post-fatigue SLD test. Importantly, increasing the knee flexion angle when landing is a known strategy to reduce injury risk (Leabeater et al., Reference Leabeater, James and Driller2022). Accordingly, the FTPW’s ability to promote a soft landing can reduce the risk of a fatigued knee injury by distributing impact forces at the knee. Consistent with this kinematic difference, we observed lower sagittal-plane loading at the knee joint, together with a corresponding reduction in ground-reaction forces. This aligns with previous research showing that adequate knee flexion significantly reduces GRF and ACL loading during high-impact activities, thereby potentially mitigating injury risk (Decker and Ziegler, Reference Decker and Ziegler2003; Kernozek et al., Reference Kernozek, Torry, van Hoof, Cowley and Tanner2005; Kulas et al., Reference Kulas, Zalewski, Hortobagyi and DeVita2008; Benjaminse et al., Reference Benjaminse, Gokeler, Dowling, Faigenbaum, Ford, Hewett and Myer2015; Beaulieu et al., Reference Beaulieu, Ashton-Miller and Wojtys2023). This finding is particularly important because women typically exhibit reduced knee flexion and higher valgus angles upon landing, factors that contribute to their higher ACL injury rates (Beutler et al., Reference Beutler, de la Motte, Marshall, Padua and Boden2009). By counteracting this tendency and enabling a deeper, soft landing, the FTPW may help lower the risk of knee injury in this population. Taken together, the FTPW’s enhancement of knee flexion and the reduction in joint loading demonstrated here in female participants directly address female-specific risk (McLean et al., Reference McLean, Walker and van den Bogert2005). The ability of FTPW to facilitate greater knee flexion while attenuating force and moment suggests its potential for injury prevention in sports and its applicability in rehabilitation, where controlled knee flexion is crucial for recovery.

In the frontal plane, the FTPW exhibited significantly greater knee varus than the PW and Ctrl during the SU test in both pre- and post-fatigue tests. Some studies have suggested that risk factors measured in the frontal plane, such as the angle and moment of knee joint valgus during landing, can lead to increased knee joint loading upon ground contact (Kar and Quesada, Reference Kar and Quesada2013; Chen et al., Reference Chen, Jiang, Yang, Cheng, Zheng and Qian2022). In addition, there is evidence that the risk of knee injury rises with a greater valgus angle (Shin et al., Reference Shin, Chaudhari and Andriacchi2009), which increases lateral knee contact pressure (Levinger et al., Reference Levinger, Menz, Morrow, Bartlett, Feller and Bergman2013). Additionally, valgus loading and internal tibial rotation coupled with valgus motion were identified as contributing factors in the ACL injury mechanism (Nessler et al., Reference Nessler, Denney and Sampley2017). By contrast, a moderate increase in varus motion can serve a protective role by distributing loads more medially and reducing excessive lateral compartment stresses, potentially mitigating ACL injury risk under certain conditions. This mechanism may be particularly relevant in activities such as the SU test, where frontal plane control is crucial for joint stability. Thus, the controlled increase in knee varus angle induced by the FTPW during the SU test may help stabilize the joint and protect it from injury, especially when muscles are fatigued and neuromuscular control is compromised.

However, it is noteworthy that excessive varus alignment of the knee can have negative implications for the prevention of osteoarthritis (OA). OA frequently manifests in the medial compartment of the knee, which experiences increased loading during knee varus (Knoop et al., Reference Knoop, Dekker, Klein, Van Der Leeden, Van Der Esch, Reiding and Steultjens2012). Subjects exhibiting varus knee patterns should not undergo excessive correction but rather be maintained within a moderate range. The application of I-taping in the lateral direction proved effective in reducing excessive varus alignment in individuals with high knee varus patterns, and this effect can be further customized based on individual needs by adjusting the taping’s stiffness.

In addition to kinematic analysis, it was possible to analyze the effects on muscle activity, metabolic cost, and cardiovascular system caused by wearing FTPW. The change in median frequency of the BF was smaller when wearing the FTPW than when wearing other garments during the fatigue protocol. As fatigue increases, the median frequency decreases, and a small change in median frequency can be interpreted as a smaller decrease in fatigue (Ament et al., Reference Ament, Bonga, Hof and Verkerke1993). Regarding muscle activity, muscle fatigue is a concern, as it can impair the muscles’ capacity to generate strength (Lattier et al., Reference Lattier, Millet, Martin and Martin2004) and can lead to altered neuromuscular control (McLean et al., Reference McLean, Fellin, Suedekum, Calabrese, Passerallo and Joy2007). These factors can subsequently induce abnormal lower limb biomechanics, thereby increasing the risk of knee injury (Santamaria and Webster, Reference Santamaria and Webster2010). In particular, the BF long head is heavily affected by fatigue, involving around 80% of all hamstring strain injuries (Bourne et al., Reference Bourne, Opar, Williams, Al Najjar and Shield2016). Although the fatigue reduction was only significant for BF, other knee flexors – ST and MGT – also showed considerable reductions in fatigue when participants wore the FTPW. Therefore, the FTPW can help injury prevention of the lower extremities by reducing fatigue in the knee flexors.

We observed a notable increase in BFR when participants wore the FTPW. This elevation in BFR aligns with findings from a previous study, where increases when using passive wearable devices were consistently noted (Palya and Kiss, Reference Palya and Kiss2020). This phenomenon is indicative of a warming effect on the lower body (Leabeater et al., Reference Leabeater, James and Driller2022). Furthermore, blood flow increases due to the compression of clothing, which reduces fatigue through oxygenation of muscles and removal of metabolites (Sperlich et al., Reference Sperlich, Born, Kaskinoro, Kalliokoski and Laaksonen2013). These factors influence the effectiveness of passive wearable devices used in sports. It is expected that it will be effective when worn for a long time, as shown in previous studies (Lee et al., Reference Lee, Sheridan, Ali, Sutanto and Wong2021).

As a result of metabolic cost analysis, there was no statistically significant difference between clothing conditions. The study implemented a fatigue protocol using the Climbmill, which necessitated the sustained maintenance of a significant knee flexion angle. Therefore, due to the reduction in fatigue among the knee flexors, such as BF, it is likely that this led to a subsequent decrease in metabolic cost. However, in line with the findings of a previous study (Lovell et al., Reference Lovell, Mason, Delphinus and McLellan2011), the wearing of passive wearable devices led to an increase in the RER, implying a potential trade-off effect between reduced fatigue and the compression effect.

The subjective assessment indicated that wearability and overall comfort did not differ between devices, whereas the FTPW elicited a stronger sensation of compression, particularly around the thigh and shank regions (Supplementary Table S1). Although participants reported this heightened pressure, the accompanying differences in comfort and tactile perception were minimal. This implies that the biomechanical benefits demonstrated by the FTPW were not compensated by additional discomfort. In addition, some of the study participants with prior experience on conventional knee braces gave subjective comments that the FTPW provided similar landing stability while allowing greater freedom of movement or comfort.

The FTPW’s elastic, body-conforming design lets athletes wear it during training or daily routines without restricting natural movement. Like standard compression garments, it is lightweight and flexible, allowing it to be incorporated into uniforms without affecting performance. By increasing knee flexion at landing, the device promotes a soft-landing strategy that lowers impact forces and injury risk in jump-intensive sports. It is also reusable, easy to apply without professional help, and less likely to irritate the skin compared to kinesio tape. These advantages make the FTPW a practical support option that offers both protection and mobility for dynamic athletic activities. However, the FTPW’s support pattern is sewn into the fabric and cannot be reoriented to suit day-to-day needs. For example, if an athlete develops quadriceps soreness and needs taping that facilitates knee extension, the current design may not offer the desired assistance. Nevertheless, for most recreational users, it still provides meaningful knee stabilization during regular exercise.

There are a few limitations in this study. The FTPW is prototyped in the laboratory, and two sizes of FTPW were used for all subjects; therefore, the intended taping patterns might not have been perfectly aligned for some subjects. Moreover, our findings are based solely on the SU and SLD tasks. Additional assessments, such as running and jump landings under various load conditions, are needed to generalize the efficacy of the FTPW to broader athletic or daily-life scenarios. Despite these limitations, the present study underscores the promise of integrating targeted elastic taping into passive wearable devices to enhance knee flexion and minimize injury risks, particularly under fatigue-induced conditions.

5. Conclusion

This study showed that the FTPW was mainly effective in promoting safer knee joint kinematics in both the frontal plane and sagittal plane. Furthermore, the FTPW exhibited biological effects resulting from knee flexion assistance and compression. These results provide an encouraging outlook for the future development of passive wearable devices for injury prevention.

Supplementary material

The supplementary material for this article can be found at https://doi.org/10.1017/wtc.2025.10022.

Data availability statement

Data can be made available to interested researchers upon request by email to the corresponding author.

Authorship contribution

S.-J.P. designed the experimental protocol, performed the experiments, and wrote the manuscript. S.C. designed and fabricated the knee flexion taping passive wearable device, performed the experiments, analyzed the results, and wrote the manuscript. H.-S.P. supervised the project and edited the manuscript.

Funding statement

This work was supported by Kolon Industries under Grant G01210007.

Competing interests

The author declares no competing interests exist OR The authors declare no competing interests exist.

Ethical standard

The authors assert that all procedures contributing to this work comply with the ethical standards of the relevant national and institutional committees on human experimentation and with the Helsinki Declaration of 1975, as revised in 2008.

Footnotes

S.-J.P. and S.C. contributed equally to this work and are designated as co-first authors.

References

Alentorn-Geli, E, Myer, GD, Silvers, HJ, Samitier, G, Romero, D, Lázaro-Haro, C and Cugat, R (2009) Prevention of non-contact anterior cruciate ligament injuries in soccer players. Part 1: Mechanisms of injury and underlying risk factors. Knee Surgery, Sports Traumatology, Arthroscopy 17, 705729.10.1007/s00167-009-0813-1CrossRefGoogle ScholarPubMed
Ament, W, Bonga, GJJ, Hof, AL and Verkerke, GJ (1993) EMG median power frequency in an exhausting exercise. Journal of Electromyography and Kinesiology 3(4), 214220.10.1016/1050-6411(93)90010-TCrossRefGoogle Scholar
Arden, N and Nevitt, MC (2006) Osteoarthritis: Epidemiology. Best Practice & Research Clinical Rheumatology 20(1), 325.10.1016/j.berh.2005.09.007CrossRefGoogle ScholarPubMed
Beaulieu, ML, Ashton-Miller, JA and Wojtys, EM (2023) Loading mechanisms of the anterior cruciate ligament. Sports Biomechanics 22(1), 129.10.1080/14763141.2021.1916578CrossRefGoogle ScholarPubMed
Benjaminse, A, Gokeler, A, Dowling, AV, Faigenbaum, A, Ford, KR, Hewett, TE and Myer, GD (2015) Optimization of the anterior cruciate ligament injury prevention paradigm: Novel feedback techniques to enhance motor learning and reduce injury risk. Journal of Orthopaedic & Sports Physical Therapy 45(3), 170182.10.2519/jospt.2015.4986CrossRefGoogle ScholarPubMed
Beutler, AI, de la Motte, SJ, Marshall, SW, Padua, DA and Boden, BP (2009) Muscle strength and qualitative jump-landing differences in male and female military cadets: The jump-ACL study. Journal of Sports Science & Medicine 8(4), 663.Google ScholarPubMed
Bourne, MN, Opar, DA, Williams, MD, Al Najjar, A and Shield, AJ (2016) Muscle activation patterns in the N ordic hamstring exercise: Impact of prior strain injury. Scandinavian Journal of Medicine & Science in Sports 26(6), 666674.10.1111/sms.12494CrossRefGoogle Scholar
Chappell, JD, Yu, B, Kirkendall, DT and Garrett, WE (2002) A comparison of knee kinetics between male and female recreational athletes in stop-jump tasks. The American Journal of Sports Medicine 30(2), 261267.10.1177/03635465020300021901CrossRefGoogle ScholarPubMed
Chen, L, Jiang, Z, Yang, C, Cheng, R, Zheng, S and Qian, J (2022) Effect of different landing actions on knee joint biomechanics of female college athletes: Based on opensim simulation. Frontiers in Bioengineering and Biotechnology 10, 899799.10.3389/fbioe.2022.899799CrossRefGoogle ScholarPubMed
Chyung, SY, Roberts, K, Swanson, I and Hankinson, A (2017) Evidence-based survey design: The use of a midpoint on the Likert scale. Performance Improvement 56(10), 1523.10.1002/pfi.21727CrossRefGoogle Scholar
Decker, NR and Ziegler, GR (2003) Mechanical properties of aerated confectionery. Journal of Texture Studies 34(4), 437448.10.1111/j.1745-4603.2003.tb01074.xCrossRefGoogle Scholar
Ford, KR, Myer, GD and Hewett, TE (2003) Valgus knee motion during landing in high school female and male basketball players. Medicine & Science in Sports & Exercise 35(10), 17451750.10.1249/01.MSS.0000089346.85744.D9CrossRefGoogle ScholarPubMed
Holcomb, WR, Rubley, MD, Lee, HJ and Guadagnoli, MA (2007) Effect of hamstring-emphasized resistance training on hamstring: Quadriceps strength ratios. The Journal of Strength & Conditioning Research 21(1), 4147.Google ScholarPubMed
Hsu, W-C, Tseng, L-W, Chen, F-C, Wang, L-C, Yang, W-W, Lin, Y-J and Liu, C (2020) Effects of compression garments on surface EMG and physiological responses during and after distance running. Journal of Sport and Health Science 9(6), 685691.10.1016/j.jshs.2017.01.001CrossRefGoogle ScholarPubMed
Joyner, MJ and Casey, DP (2015) Regulation of increased blood flow (hyperemia) to muscles during exercise: A hierarchy of competing physiological needs. Physiological Reviews 95(2), 549601.10.1152/physrev.00035.2013CrossRefGoogle ScholarPubMed
Kadaba, MP, Ramakrishnan, HK and Wootten, ME (1990) Measurement of lower extremity kinematics during level walking. Journal of Orthopaedic Research 8(3), 383392.10.1002/jor.1100080310CrossRefGoogle ScholarPubMed
Kar, J and Quesada, PM (2013) A musculoskeletal modeling approach for estimating anterior cruciate ligament strains and knee anterior–posterior shear forces in stop-jumps performed by young recreational female athletes. Annals of Biomedical Engineering 41, 338348.10.1007/s10439-012-0644-yCrossRefGoogle ScholarPubMed
Kernozek, TW, Torry, MR, van Hoof, H, Cowley, H and Tanner, S (2005) Gender differences in frontal and sagittal plane biomechanics during drop landings. Medicine & Science in Sports & Exercise 37(6), 10031012.Google ScholarPubMed
Knoop, J, Dekker, J, Klein, J-P, Van Der Leeden, M, Van Der Esch, M, Reiding, D and Steultjens, MPM (2012) Biomechanical factors and physical examination findings in osteoarthritis of the knee: Associations with tissue abnormalities assessed by conventional radiography and high-resolution 3.0 Tesla magnetic resonance imaging. Arthritis Research & Therapy 14, 112.10.1186/ar4050CrossRefGoogle ScholarPubMed
Kramer, A, Kümmel, J, Gollhofer, A, Armbrecht, G, Ritzmann, R, Belavy, D and Gruber, M (2018) Plyometrics can preserve peak power during 2 months of physical inactivity: An RCT including a one-year follow-up. Frontiers in Physiology 9, 382250.10.3389/fphys.2018.00633CrossRefGoogle ScholarPubMed
Kulas, A, Zalewski, P, Hortobagyi, T and DeVita, P (2008) Effects of added trunk load and corresponding trunk position adaptations on lower extremity biomechanics during drop-landings. Journal of Biomechanics 41(1), 180185.10.1016/j.jbiomech.2007.06.027CrossRefGoogle ScholarPubMed
Lattier, G, Millet, GY, Martin, A and Martin, V (2004) Fatigue and recovery after high-intensity exercise part II: Recovery interventions. International Journal of Sports Medicine 25(7), 509515.10.1055/s-2004-820946CrossRefGoogle ScholarPubMed
Laughlin, WA, Weinhandl, JT, Kernozek, TW, Cobb, SC, Keenan, KG and O’Connor, KM (2011) The effects of single-leg landing technique on ACL loading. Journal of Biomechanics 44(10), 18451851.10.1016/j.jbiomech.2011.04.010CrossRefGoogle ScholarPubMed
Leabeater, AJ, James, LP and Driller, MW (2022) Tight margins: Compression garment use during exercise and recovery – A systematic review. Textiles 2(3), 395421.10.3390/textiles2030022CrossRefGoogle Scholar
Lee, H, Kim, SH and Park, H-S (2020) A fully soft and passive assistive device to lower the metabolic cost of sit-to-stand. Frontiers in Bioengineering and Biotechnology 8, 966.10.3389/fbioe.2020.00966CrossRefGoogle ScholarPubMed
Lee, DCW, Sheridan, S, Ali, A, Sutanto, D and Wong, SHS (2021) Wearing compression tights post-exercise enhances recovery hemodynamics and subsequent cycling performance. European Journal of Applied Physiology 121, 20912100.10.1007/s00421-021-04661-0CrossRefGoogle ScholarPubMed
Levinger, P, Menz, HB, Morrow, AD, Bartlett, JR, Feller, JA and Bergman, NR (2013) Relationship between foot function and medial knee joint loading in people with medial compartment knee osteoarthritis. Journal of Foot and Ankle Research 6, 19.10.1186/1757-1146-6-33CrossRefGoogle ScholarPubMed
Lovell, DI, Mason, DG, Delphinus, EM and McLellan, CP (2011) Do compression garments enhance the active recovery process after high-intensity running? The Journal of Strength & Conditioning Research 25(12), 32643268.10.1519/JSC.0b013e31821764f8CrossRefGoogle ScholarPubMed
McLean, SG, Walker, KB and van den Bogert, AJ (2005) Effect of gender on lower extremity kinematics during rapid direction changes: An integrated analysis of three sports movements. Journal of Science and Medicine in Sport 8(4), 411422.10.1016/S1440-2440(05)80056-8CrossRefGoogle ScholarPubMed
McLean, SG, Fellin, RE, Suedekum, N, Calabrese, G, Passerallo, A and Joy, S (2007) Impact of fatigue on gender-based high-risk landing strategies. Medicine and Science in Sports and Exercise 39(3), 502514.10.1249/mss.0b013e3180d47f0CrossRefGoogle ScholarPubMed
McNeil, CJ, Allen, MD, Olympico, E, Shoemaker, JK and Rice, CL (2015) Blood flow and muscle oxygenation during low, moderate, and maximal sustained isometric contractions. American Journal of Physiology-Regulatory, Integrative and Comparative Physiology 309(5), R475R481.10.1152/ajpregu.00387.2014CrossRefGoogle ScholarPubMed
Mj, K (1957) The effects of training on heart rate: A longitudinal study. Annales Medicinae Experimentalis et Biologiae Fenniae 35, 307315.Google Scholar
Myer, GD, Paterno, MV, Ford, KR, Quatman, CE and Hewett, TE (2006) Rehabilitation after anterior cruciate ligament reconstruction: Criteria-based progression through the return-to-sport phase. Journal of Orthopaedic & Sports Physical Therapy 36(6), 385402.10.2519/jospt.2006.2222CrossRefGoogle ScholarPubMed
Nessler, T, Denney, L and Sampley, J (2017) ACL injury prevention: What does research tell us? Current Reviews in Musculoskeletal Medicine 10, 281288.10.1007/s12178-017-9416-5CrossRefGoogle ScholarPubMed
Palya, Z and Kiss, RM (2020) Biomechanical analysis of the effect of compression sportswear on running. Materials Today: Proceedings 32, 133138.Google Scholar
Pataky, TC, Robinson, MA and Vanrenterghem, J (2013) Vector field statistical analysis of kinematic and force trajectories. Journal of Biomechanics 46(14), 23942401.10.1016/j.jbiomech.2013.07.031CrossRefGoogle Scholar
Paz, GA, de Freitas Maia, M, Farias, D, Santana, H, Miranda, H, Lima, V and Herrington, L (2016) Kinematic analysis of knee valgus during drop vertical jump and forward step-up in young basketball players. International Journal of Sports Physical Therapy 11(2), 212.Google ScholarPubMed
Ramsey, DK, Briem, K, Axe, MJ and Snyder-Mackler, L (2007) A mechanical hypothesis for the effectiveness of knee bracing for medial compartment knee osteoarthritis. The Journal of Bone and Joint Surgery American 2007, 89.Google Scholar
Rishiraj, N, Taunton, JE, Lloyd-Smith, R, Woollard, R, Regan, W and Clement, DB (2009) The potential role of prophylactic/functional knee bracing in preventing knee ligament injury. Sports Medicine 39, 937960.10.2165/11317790-000000000-00000CrossRefGoogle ScholarPubMed
Santamaria, LJ and Webster, KE (2010) The effect of fatigue on lower-limb biomechanics during single-limb landings: A systematic review. Journal of Orthopaedic & Sports Physical Therapy 40(8), 464473.10.2519/jospt.2010.3295CrossRefGoogle ScholarPubMed
Schorer, R, Badoual, A, Bastide, B, Vandebrouck, A, Licker, M and Sage, D (2017) A feasibility study of color flow doppler vectorization for automated blood flow monitoring. Journal of Clinical Monitoring and Computing 31, 11671175.10.1007/s10877-016-9953-2CrossRefGoogle ScholarPubMed
Shin, CS, Chaudhari, AM and Andriacchi, TP (2009) The effect of isolated valgus moments on ACL strain during single-leg landing: A simulation study. Journal of Biomechanics 42(3), 280285.10.1016/j.jbiomech.2008.10.031CrossRefGoogle ScholarPubMed
Sperlich, B, Born, D-P, Kaskinoro, K, Kalliokoski, KK and Laaksonen, MS (2013) Squeezing the muscle: Compression clothing and muscle metabolism during recovery from high intensity exercise. PLoS One 8(4), e60923.10.1371/journal.pone.0060923CrossRefGoogle ScholarPubMed
Troynikov, O, Ashayeri, E, Burton, M, Subic, A, Alam, F and Marteau, S (2010) Factors influencing the effectiveness of compression garments used in sports. Procedia Engineering 2(2), 28232829.Google Scholar
Wang, Y, Liu, H, Wei, H, Wu, C and Yuan, F (2023) The effect of a knee brace on muscle forces during single-leg landings at two heights. International Journal of Environmental Research and Public Health 20(5), 4652.10.3390/ijerph20054652CrossRefGoogle ScholarPubMed
Wu, G, Siegler, S, Allard, P, Kirtley, C, Leardini, A, Rosenbaum, D and Witte, H (2002) ISB recommendation on definitions of joint coordinate system of various joints for the reporting of human joint motion – Part I: Ankle, hip, and spine. Journal of Biomechanics 35(4), 543548.10.1016/S0021-9290(01)00222-6CrossRefGoogle ScholarPubMed
Yang, C-S and Lim, B-O (2014) Effects of knee brace on the anterior cruciate ligament injury risk factors during spike take off in female volleyball players. Korean Journal of Sport Biomechanics 24(1), 2733.10.5103/KJSB.2014.24.1.027CrossRefGoogle Scholar
Figure 0

Figure 1. Design overview of FTPW. X-taping for knee flexion was applied to the back of the knee, and I-taping for preventing knee abduction/adduction was applied to the medial/lateral sides of the knee, respectively. To prevent slipping during motion, bands were used to anchor the garment to the ankle and waist.

Figure 1

Figure 2. The experimental procedure in this study. All participants completed three trials, at a 1-week interval. SLD/SU tests were performed before and after fatigue riding the Climbmill.

Figure 2

Figure 3. Knee joint rotation during the SLD/SU test (a, b/c, d) before the fatigue protocol. The black, blue, and magenta lines represent Ctrl, PW, and FTPW, respectively. Error bars mean standard error. The gray-shaded area corresponds to sections where the repeated-measures ANOVA results indicate a statistically significant difference (p < .05). Below each graph, bars compare group differences at each time point: Ctrl versus PW, Ctrl versus FTPW, and PW versus FTPW. The color of bars corresponds to the group with larger values.

Figure 3

Figure 4. Knee joint rotation during the SLD/SU test (a, b/c, d) after the fatigue protocol. The black, blue, and magenta lines represent Ctrl, PW, and FTPW, respectively. Error bars mean standard error. The gray-shaded area corresponds to sections where the repeated-measures ANOVA results indicate a statistically significant difference (p < .05). Below each graph, bars compare group differences at each time point: Ctrl versus PW, Ctrl versus FTPW, and PW versus FTPW. The color of bars corresponds to the group with larger values.

Figure 4

Figure 5. The median frequency difference of each muscle during the first and last 30 s of Climbmill exercise for the different conditions. Error bars mean standard error, and the value above the error bars is mean ± standard error. Asterisks (* and **) indicate statistically significant differences between clothing conditions (p < .05 and p < .01).

Figure 5

Figure 6. The RER (a) and BFR (b) for each condition. Error bars mean standard error, and the value above the error bars is mean ± standard error. Asterisk (*) indicates statistically significant differences between clothing conditions (p < .05).